Print
Category: News
Hits: 1804

Smoothing of the Ablated Porcine Anterior Corneal Surface Using the Technolas Keracor 217C and Nidek EC-5000 Excimer Lasers

Lombardo M, Serrao S
J Refract Surg. 2004 Sep-Oct;20(5):450-3.
PMID: 15523956

Abstract

PURPOSE:

To demonstrate efficacy of a smoothing technique to increase regularity of the anterior corneal surface after photorefractive keratectomy (PRK), using two different excimer lasers.

METHODS:

Spherical ablations of -10.00 D were performed on 11 fresh porcine corneas using either the Technolas Keracor 217C scanning-spot or the Nidek EC-5000 scanning-slit beam excimer laser. Following the procedure, we performed a phototherapeutic keratectomy treatment (smoothing technique) on half of the corneal surface. The smoothing technique was performed using a viscous solution of 0.25% sodium hyaluronate, which was spread on the cornea prior to the procedure. The ablation zone was 6 mm in diameter and the transition zone extended to 3 mm. The ablation depth was set at 10 microm. Corneas were then examined with scanning electron microscopy.

RESULTS:

Smoother treatment zones were apparent in porcine corneas in which smoothing was performed following PRK, with both laser systems. Results from the two lasers were not directly compared.

CONCLUSIONS:

The smoothing procedure performed following PRK using a viscous 0.25% sodium hyaluronate masking solution and a scanning laser system rendered the porcine corneal surface more regular.